MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS; DEVICE, IRRIGATION, OCULAR SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS; DEVICE, IRRIGATION, OCULAR SURGERY Back to Search Results
Catalog Number 8065751511
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that before the cataract surgery, polymer tip part was found detached from the handpiece.Surgery was completed by replacing alternate product.Patient impact was not reported.
 
Manufacturer Narrative
The returned samples were visually inspected and deemed non-conforming, with the over mold tip missing from the end of the metal cannula.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.The returned sample was found to be non-conforming visually the tip of the polymer part was found detached, as described in the complaint.How and when the polymer tip became detached cannot be confirmed and a root cause for the detached tip cannot be determined from the evaluation performed; however, a manufacturing related issue cannot be ruled out.A non-conformance investigation has been initiated in order to investigate the root cause for the polymer tip detachment.All i/a tips are 100% visually inspected prior to being released from the manufacturing facility.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS
Type of Device
DEVICE, IRRIGATION, OCULAR SURGERY
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18509671
MDR Text Key332989460
Report Number2523835-2024-00029
Device Sequence Number1
Product Code KYG
UDI-Device Identifier00380657515110
UDI-Public00380657515110
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751511
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-