ALCON RESEARCH, LLC - ALCON PRECISION DEVICE DISPOSABLE POLYMER IRRIGATION AND SILICONE IRRIGATION AND ASPIRATION TIPS; DEVICE, IRRIGATION, OCULAR SURGERY
|
Back to Search Results |
|
Catalog Number 8065751511 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A physician reported that before the cataract surgery, polymer tip part was found detached from the handpiece.Surgery was completed by replacing alternate product.Patient impact was not reported.
|
|
Manufacturer Narrative
|
The returned samples were visually inspected and deemed non-conforming, with the over mold tip missing from the end of the metal cannula.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.The returned sample was found to be non-conforming visually the tip of the polymer part was found detached, as described in the complaint.How and when the polymer tip became detached cannot be confirmed and a root cause for the detached tip cannot be determined from the evaluation performed; however, a manufacturing related issue cannot be ruled out.A non-conformance investigation has been initiated in order to investigate the root cause for the polymer tip detachment.All i/a tips are 100% visually inspected prior to being released from the manufacturing facility.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.The manufacturer internal reference number is: (b)(4).
|
|