Model Number 0998-00-0800-65
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) unit had frequent "iab circuit leakage (gas loss)" alarms.A possible balloon rupture.Patient's condition stable and iab withdrawn.
 
Manufacturer Narrative
No further information available.A supplemental report will be sent if additional information becomes available.Not returned to manufacturer.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and was not able to confirm the reported issue.No blood flowed into the equipment.All manifold tests passed.No problems were found with the equipment, so a problem with the balloon was suspected.Each regulator was adjusted and a running test was performed and it worked well.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18510863
MDR Text Key332916955
Report Number2249723-2024-00135
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-65
Device Catalogue Number0998-00-0800-65
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/17/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/19/2023
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-