• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER, 230 V CONTINENTAL EUROPE; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER, 230 V CONTINENTAL EUROPE; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Overheating of Device (1437)
Patient Problem Insufficient Information (4580)
Event Date 12/01/2023
Event Type  malfunction  
Manufacturer Narrative
B3: date of event is unknown, no information has been provided to date.Device evaluation: one device was returned for investigation.Visual inspection found the unit looks in used condition.No cracks or leaks were visible.Functional testing could not duplicate or confirm the complaint.The device heats the circulation solution to 42.0 degrees and remains stable.Unit runs for more than 2 hours with no temperature variations detected.Checked all alarm settings, no problem found.Preventive maintenance performed.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Event Description
It was reported that the temperature was greater than 90 degrees.Patient involvement is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER, 230 V CONTINENTAL EUROPE
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18511103
MDR Text Key333354681
Report Number2183161-2024-00045
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public(01)50695085820909(11)090617
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberHL-90-INT-230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2023
Date Manufacturer Received12/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-