Model Number HL-90
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2023
Event Type  malfunction  
Event Description
It was reported that the device did not power on.The reporter stated they did not have any information other than it would not turn on.They "opened it up to look at it and found out that one the prongs from the power button broke from the board inside.He soldered it back together and it now will power on but, now it alarms high temperature." no patient involvement or clinical harm was reported.
 
Manufacturer Narrative
B3: date of event is unknown, no information has been provided to date.E1 initial reporter additional phone number: (b)(6).Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
One device was received with multiple attachments on the front cover, microswitch was bent, pole clamp discolored, line cord faded and worn, corroded quick connect, enclosure cracked under quick connect, water tank cover cracked on all four corners, outdated printed circuit board (pcb) and power source.Per functional testing, the device did not turn on.The complaint was confirmed.The root cause was pcb.It was unknown what caused the condition.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18511241
MDR Text Key332882009
Report Number2183161-2024-00050
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public(01)50695085820909(11)091201
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/02/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received01/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-