Model Number HL-90
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/16/2023
Event Type  malfunction  
Event Description
It was reported that after switching on the device, a noise could be heard even on a "flat plane".No patient involvement was reported.
 
Manufacturer Narrative
E1 initial reporter phone number: (b)(6).Device evaluation: one device was received in new condition, no cracks, leaks, or damages were visible.Per functional testing, an unusual operating noise could be heard after switching on.The complaint was confirmed.The root cause was a faulty water pump.It was unknown what caused the condition.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the pump and performed preventative maintenance (pm).The device passed all functional and delivery tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTINENTAL EUROPEAN 230 VOLT LEVEL 1 HOTLINE
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18511819
MDR Text Key333330641
Report Number2183161-2024-00051
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberHL-90-INT-230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/08/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse