• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CAPSURE EPICARDIAL LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CAPSURE EPICARDIAL LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 4968-60
Device Problems Mechanical Problem (1384); Pacing Problem (1439)
Patient Problem Cardiac Arrest (1762)
Event Date 01/05/2024
Event Type  Death  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that post operative the patient arrested post implant of an implantable pulse generator (ipg) and pacing lead. it was reported that the ipg system exhibited no pacing.It was reported that battery depletion was noted.Resuscitation and temporary pacing was used.The pocket was reopened pocket and the ipg and lead were disconnected and reconnected several times without success.The ipg and lead were eventually connected and it was reported that the system was not functional.The ipg was explanted and replaced with another ipg. it was reported that the patient died approximately three days later.Cause of death was provided as kidney failure and ventricular septal defect.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was further reported that the pacing lead was implanted in the right ventricle (rv).It was confirmed that the patient's death was not associated with the performance of the ipg or rv lead.It was also noted that the no ipg premature battery depletion was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE EPICARDIAL LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18516296
MDR Text Key332868094
Report Number2182208-2024-00163
Device Sequence Number1
Product Code DTB
Combination Product (y/n)Y
Reporter Country CodeTH
PMA/PMN Number
P950024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number4968-60
Device Catalogue Number4968-60
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
W3SR01 IPG, W3SR01 IPG.
Patient Outcome(s) Death;
Patient Age12 MO
Patient SexFemale
Patient Weight6 KG
-
-