ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number D-AVHD-DF16 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/21/2023 |
Event Type
malfunction
|
Event Description
|
Related manufacturing reference: 3008452825-2024-00029, 3008452825-2024-00030, 3008452825-2024-00031.During the atrial fibrillation procedure, communication issues required multiple troubleshooting measures and resulted in a procedural delay.The devices were all exchanged and the procedure was completed with no adverse consequences to the patient.The patches were applied to the patient, the ablation catheter and the mapping catheter were inserted into the heart, mapping was performed, rf delivery was started and the right pulmonary vein isolation was started.When the right pulmonary vein isolation was performed to the middle, it was noted that the electrodes 2-4 of the ablation catheter were displayed in red and the navigation became unable to be displayed.The direct connect cable for ablation catheter extension was exchanged with no resolution.The ablation catheter was replaced with no resolution.The ensite was restarted and resume study was performed, but the same issue occurred with both the catheters.The issue was resolved after registering as a new patient and replacing the patches, replacing the ablation catheter, and replacing the mapping catheter at the same time.The procedure was completed with no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Additional information: d9, g3, h2, h3, h6 one bi-directional, curve d-f, sensor enabled, advisor hd grid mapping catheter was received for evaluation.Electrodes 17-18 and the sensors met specifications for acceptable resistance values with no open or short circuits detected.Additionally, the catheter was successfully connected to and recognized by an ensite x system.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported communication issue remains unknown.
|
|
Search Alerts/Recalls
|
|
|