COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIG60AXT
Device Problems Entrapment of Device (1212); Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  malfunction  
Event Description
According to the reporter, during the laparoscopic thoracoscopic lobectomy, the thick pulmonary parenchyma was resected, on midway, though heavy, the handle managed to fire until the end.It was attempted to return the black return knob, but it was extremely hard. several attempts were made and the black return knob managed to return.It was discovered that the jaws could not be opened.The reload was then removed from the tissue when pulled while moving the hand little by little.There was no patient injury.
 
Manufacturer Narrative
D10 concomitant product: sig60axt - tri 2.0 sig60axt 60 xtra thk 15mm, lot# unknown medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the reload was fully fired, the jaws of the reload were open, and the clamping mechanism was deformed.It was reported that the instrument locked on tissue and it was difficult to retract the knife blade.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The evaluation detected unreported conditions: damaged knife blade, thick tissue and lock ring out of position.Visual inspection noted the reload knife laminates were damaged, the reload had damage to the cutting edge of the knife blade and lock ring was in the home position.The product analysis noted evidence that the device was not used as intended.This issue may occur if the lock ring is inadvertently moved out of position during handling of the reload.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: always inspect the tissue thickness and select an appropriate staple size prior to application of the endo gia¿ universal stapler.When positioning the stapler on the application site, ensure that no obstructions, such as clips, are incorporated into the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18519938
MDR Text Key333233253
Report Number1219930-2024-00289
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884521543591
UDI-Public10884521543591
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K160176
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSIG60AXT
Device Catalogue NumberSIG60AXT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10
-
-