|
Model Number EGIAUSHORT |
Device Problems
Entrapment of Device (1212); Retraction Problem (1536)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/21/2023 |
Event Type
malfunction
|
Event Description
|
According to the reporter, during the laparoscopic thoracoscopic lobectomy, the thick pulmonary parenchyma was resected, on midway, though heavy, the handle managed to fire until the end.It was attempted to return the black return knob, but it was extremely hard. several attempts were made and the black return knob managed to return.It was discovered that the jaws could not be opened.The reload was then removed from the tissue when pulled while moving the hand little by little.There was no patient injury.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: d9 (latest return date), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted instrument firing knobs were retracted and the articulation lever was in neutral position.It was reported that the instrument locked on tissue and it was difficult to retract the knife blade.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The evaluation detected an unreported condition: sheared front teeth.Visual inspection noted that there was an observation hole on the handle for the firing rack and sheared teeth were visible on the firing rack at site of initial firing post clamp up.Functionally the instrument loaded, clamped, yet would not cycle due to the sheared firing rack teeth damage.The product analysis noted evidence that the device was not used as intended.This issue rack may occur in any of the following circumstances: firing over tissue that is beyond the recommended thickness range; firing with an obstacle incorporated in the jaws; attempting to fire a reload that is in interlock.In each circumstance, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: ensure that the black return knob on the instrument is pulled back completely and the articulation lever is neutral to the instrument.Cycle the instrument to confirm proper loading of the reload.To do so, squeeze the handle once to close the jaws of the reload.Push the black loop handle all of the way forward or pull back on the black return knob and confirm that the reload jaws open fully.Preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|