Catalog Number SGC0701
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Air Embolism (1697); Non specific EKG/ECG Changes (1817)
Event Date 01/04/2024
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The investigation determined the reported air embolism appears to be related to user technique.The st elevation appears to be a cascading effect of the air embolism.Air embolism and st elevation (non specific ekg/ecg changes) are listed in the instructions for use as known possible complications associated with the mitraclip procedures.The reported serious injury/illness/impairment was a result of case specific circumstance as no intervention was performed.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that on (b)(6) 2024, a mitraclip procedure was performed to treat degenerative mitral regurgitation grade 4+.The nurse removed the dilator, but it was not coaxial so air was introduced into the patient.The air was visualized on fluoroscopy.St elevation was noted.After five minutes, the st elevation subsided spontaneously.No intervention was performed.The clip was implanted successfully, reducing mr to grade 1+.There was no clinically significant delay.The patient is discharged and doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP¿ G4 STEERABLE GUIDE CATHETER (MDR)
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18520409
MDR Text Key332947959
Report Number2135147-2024-00262
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberSGC0701
Device Lot Number31031R2107
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age73 YR
Patient SexMale
Patient Weight79 KG
-
-