MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CA500, EPIX UNIVERSAL CLIP APPLIER 3/BX; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CA500, EPIX UNIVERSAL CLIP APPLIER 3/BX; CLIP, IMPLANTABLE Back to Search Results
Model Number CA500
Device Problem Failure to Fire (2610)
Patient Problem Laceration(s) (1946)
Event Date 12/19/2023
Event Type  Injury  
Manufacturer Narrative
The event unit is anticipated to return to applied medical.A follow up report will be provided following the completion of the investigation.
 
Event Description
Procedure performed: cholecstectomy.Event description: complaint 1 of 2: (b)(4).Complaint 2 of 2: (b)(4).Translation: applies to applied's ca500 clip-on pliers.The clip does not release the clips each time the handle is activated.Resulting in tissue shearing and untimely bleeding.Other pliers with the same problem.Additional information received from applied medical representative via complaint form on 20dec2023: clip of cystic artery, no clip delivered and section of cystic artey, bleeding.Additional information received from applied medical representative via email on 29dec23: the trigger could be moved as normal.A clip did not seat into the jaws properly.The applicator jaws tore the tissue.Patient status: ok.Intervention: device replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CA500, EPIX UNIVERSAL CLIP APPLIER 3/BX
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
farah azmi
22872 avenida empresa
rancho santa margarita, CA 92688
9497138710
MDR Report Key18522941
MDR Text Key332974162
Report Number2027111-2024-00338
Device Sequence Number1
Product Code FZP
UDI-Device Identifier00607915125318
UDI-Public(01)00607915125318(17)260503(30)01(10)1490524
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K011236
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCA500
Device Catalogue Number101474072
Device Lot Number1490524
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/20/2023
Date Device Manufactured05/04/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-