C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 1606052 |
Device Problems
Fracture (1260); Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a port placement procedure, the syringe was allegedly broken and could not be used.The procedure was completed by using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one monoject syringe was returned for evaluation.Visual evaluation was performed on the returned device.Visual evaluation shows the rubber seal inside the syringe was partially detached from the plunger.Therefore the investigation is confirmed for the identified detachment issue.However the investigation is unconfirmed for the reported material integrity issue as the more specific damage such as detachment is identified during investigation.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: g3 h11: b5, h6 (device, method, result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
|
Event Description
|
It was reported that during a port placement procedure, the syringe was allegedly damaged and could not be used.The procedure was completed by using another device.There was no reported patient injury.
|
|