Model Number 0998-00-0800-52
Device Problem Unexpected Shutdown (4019)
Patient Problems Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2023
Event Type  Death  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient having a cabg procedure, the cardiosave intra-aortic balloon pump (iabp) shutdown during operational procedure.The patient expired.
 
Manufacturer Narrative
Updated fields: b4, d9, g3, g6, h2, h3, h4, h6(health effect ¿ clinical.Type of investigation, investigation findings, investigation conclusions), h10.A getinge field service engineer (fse) was not dispatched to evaluate the iabp unit because the customer stated the machine was plugged-in the whole time prior wheeling patient out of the operating room, the batteries were completely charged prior to transport.Upon engineer check-up the coil upon the monitor was loose and the cause of the machine shutdown.The customer already replaced the coiled cord with a new one, and its already functional.
 
Event Description
N/a.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE G PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18525221
MDR Text Key333026230
Report Number2249723-2024-00192
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108438
UDI-Public10607567108438
Combination Product (y/n)N
Reporter Country CodeRP
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-52
Device Catalogue Number0998-00-0800-52
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age71 YR
Patient SexFemale
Patient Weight46 KG
-
-