• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ORG-9110A; MULTIPLE PATIENT RECEIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ORG-9110A; MULTIPLE PATIENT RECEIVER Back to Search Results
Model Number ORG-9110A
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Manufacturer Narrative
The customer reported that this facility is having tele communication dropping out for 30 seconds.The teles go into communication loss (comm loss) for all the beds then they go into bed not found and then come out of comm loss.Technical support (ts) asked the customer to check the org closet to see if the org's can be rebooted and check for error lights or any error lights on the switches.The customer stated he didn't see any error lights, but turned off the orgs for 30 seconds and rebooted the central nurse's station (cns); however, this did not resolve the issue.The customer will send in the unit to be repaired.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.D10 concomitant medical device: the following device was used in conjunction with the org: cns: model #: ni.Serial #: ni.Device manufacturer date: ni unique identifier (udi) #: ni returned to nihon kohden: ni.D10 concomitant medical device: the following device was used in conjunction with the org: zm transmitters: model #: ni serial #: ni device manufacturer date: ni unique identifier (udi) #: ni returned to nihon kohden: ni.
 
Event Description
The customer reported that this facility is having tele communication dropping out for 30 seconds.There was no patient injuryreported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORG-9110A
Type of Device
MULTIPLE PATIENT RECEIVER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key18527297
MDR Text Key333478239
Report Number8030229-2024-04017
Device Sequence Number1
Product Code DRG
UDI-Device Identifier04931921103906
UDI-Public04931921103906
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORG-9110A
Device Catalogue NumberORG-9110A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CNS; ZM TRANSMITTERS
-
-