• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number FRX500S14
Device Problems Corroded (1131); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  malfunction  
Manufacturer Narrative
H3 other text : device serviced by a third-party service center.
 
Event Description
The manufacturer received information regarding a dreamstation auto device.The device was returned to a third-party service center.During visual inspection of the device, corrosion was found on the power connector and on metallic surface of the device.In addition, the inspection found unit can't be power on, contamination from dust/dirt and the device was scrapped.This report is being submitted for the corrosion was found on the power connector and on metallic surface of the device during service.There was no allegation of serious or permanent harm or injury.No medical intervention was required for the patient.
 
Manufacturer Narrative
The manufacturer received information in relation to a dreamstation auto unit.The device was returned to a third party service center.During visual inspection of the device, it was determined the power connector of the unit was corroded.Device was scrapped at customer's request.Analysis of past events has not indicated an adverse event has occurred due to corrosion.Review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.Additionally, the risk file indicates that corrosion will not substantially affect the performance of the device.This complaint is considered not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18527814
MDR Text Key333050739
Report Number2518422-2024-02996
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberFRX500S14
Device Catalogue NumberFRX500S14
Was Device Available for Evaluation? No
Date Manufacturer Received12/21/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/04/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-