It was reported that unspecified bd infusion set tubing expands the following information was received by the initial reporter with the following verbatim: typically, the issue occurs when excessive pressures are allowed to reach the pumping chamber.E.G., if someone administers an iv push but does not clamp above the smartsite.The attached tip sheet is not in circulation anymore, but for your reference, it talks about this in the faq section.
|
The customer reported that there was a bulge in the silicone segment.One photo was taken by the customer that verifies the complaint.Although a sample was not received, the probable root cause for the failure is determined to be a user issue.If the infusion set is not loaded correctly, a ballooning in the pumping segment tubing is possible.The clinical team has been made aware of the issue and the issue has been communicated with the customer with additional instructional material for proper loading of the alaris pumps.A device history record review could not be performed because a model and lot number was not provided by the customer.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
|