• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 86 CM; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 86 CM; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 1457Q/86
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the left ventricular (lv) lead was failing to advance over the guidewire.The lv lead was not used.The physician continued with second lv lead and completed the procedure.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of guidewire failure to advance was confirmed.A complete lead was returned in one piece for analysis.Electrical testing was normal with no anomalies found.Visual inspection found the inner coil to be kinked with excessive blood at the double bent region of the lead.The cause of the reported event was due to the inner coil being damaged consistent with occurring at the time of procedural.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD DOUBLE BEND, 86 CM
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18532131
MDR Text Key333088845
Report Number2017865-2024-01409
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734510189
UDI-Public05414734510189
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1457Q/86
Device Lot NumberA000136919
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA TACHY MRI LEAD; TENDRIL LEAD
Patient Age73 YR
Patient SexFemale
-
-