• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NATUS MEDICAL INCORPORATED IP PTZ ERGOJUST VIDEO EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NATUS MEDICAL INCORPORATED IP PTZ ERGOJUST VIDEO EXTENSION Back to Search Results
Model Number IP-EC-VIDEO-EXT
Device Problems Crack (1135); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/02/2024
Event Type  malfunction  
Manufacturer Narrative
Initial report ref natus complaint(b)(4).Technologist was rolling the machine down the hallway and noticed an extreme tilt of the camera tower.She brought it back to the lab where they investigated the tower to figure out why it was leaning.They then observed cracks at the base of the tower.The remaining qty of equipment was examined and it was noted cracks in 6 of 7 machines purchased between (b)(6).(b)(6).Rev 69 xltek eeg psg risk analysis spreadsheet identifies hazard 6.11 - due to mechanical stress or impact, dynamic loading, vibration, or environmental conditions -- accessory shelf on cart may be unable to support the mass of accessories or items placed on its surface, causing it to bend or break and leave sharp edges exposed or broken support system could land on and harm someone.Effect (harm): range from clinically insignificant to irreversible injury resulting from direct physical injury.The hazards identified have been evaluated and found to be in compliance with a known standard.As noted in (b)(6).Hazards that have been evaluated and found to be in compliance with known standards can be presumed to be consistent with an acceptable level of risk, yielding an acceptable risk / benefit to the patient or user.Risk outweighed by benefit of use of device.Install date:(b)(6).2021 pn (b)(4).-ergojust bracket kits were sent to the customer.This kit applied to the cart will stop the video extension from falling over.Field service have installed the brackets on all of the carts, all carts are now back in use.Failure confirmed: yes resolution: sending of ergojust bracket kits resolution code: preventative maintenance customer symptom code: broken/damaged part investigation case closed (b)(6).2024.
 
Event Description
Ip ptz ergojust video extension -video extensions on 6 of 7 cart purchased in 2021 have cracks.Camera arms are breaking off.No injuries.Event: 5/6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IP PTZ ERGOJUST VIDEO EXTENSION
Type of Device
IP PTZ ERGOJUST VIDEO EXTENSION
Manufacturer (Section D)
NATUS MEDICAL INCORPORATED
dba excel-tech ltd.
2568 bristol circle
oakville, ontarios, L6H 5 S1
CA  L6H 5S1
Manufacturer (Section G)
NATUS MANUFACTURING LIMITED
ida business park
gort
galway h91pd92,
EI  
Manufacturer Contact
gráinne walsh
dba excel-tech ltd.
2568 bristol circle
oakville, ontarios, L6H 5-S1
CA   L6H 5S1
MDR Report Key18533341
MDR Text Key333091968
Report Number9612330-2024-00008
Device Sequence Number1
Product Code OMB
UDI-Device Identifier00382830058377
UDI-Public00382830058377
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIP-EC-VIDEO-EXT
Device Catalogue NumberIP-EC-VIDEO-EXT
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/03/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-