• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A Back to Search Results
Catalog Number SL-2000M2095
Device Problems Leak/Splash (1354); Material Puncture/Hole (1504); Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
 
Event Description
As reported by the user facility: brief inquiry description.Cuts in arterial pod end of blood pump segment.Detailed inquiry description.Right at the initiating of hd tx staff noticed that micro air bubbles on the top of dialyzer header and in venous chamber, upon inspection clinical practitioner found a few drops of blood at the blood pump roller track.Treatment was terminated.Upon inspection of the blood line, we found tiny pinched/cut like damage right above the arterial pod of the bloodline.Defected bloodline was rinsed with bleach and water.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).A device history record (dhr) review was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.One (1) photo was submitted to the manufacturer for evaluation.Through visual examination of the photo, it was observed that the tubing had a small cut.The sample within the photo is not with specification; the defect of tubing cut is confirmed.While the defect is confirmed, an exact root cause could not be determined.During the assembly/packaging process, some type of device, knife or razor that could generate a cut in the material is not handled, and the 100% leak test is carried out on the material during the manufacturing process.A test was carried out to verify the detection of leakage in the leak tester and the results show that the machine rejects the set if the product has a cut, leak in the pump segment or any other component.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMODIALYSIS BLOODLINES
Type of Device
SET, TUBING, BLOOD, WITH A
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th avenue
bethlehem 18018
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th avenue
bethlehem PA 18018 0027
Manufacturer Contact
jonathan severino
901 marcon blvd
allentown, PA 18109
4847197287
MDR Report Key18535891
MDR Text Key333148276
Report Number2521402-2024-00008
Device Sequence Number1
Product Code FJK
UDI-Device Identifier04046964367762
UDI-Public(01)04046964367762
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSL-2000M2095
Device Lot NumberA2300208
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-