ABBOTT MEDICAL AMPLATZER VASCULAR PLUG II; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Catalog Number 9-AVP2-012
Device Problem Difficult to Fold, Unfold or Collapse (1254)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2023
Event Type  malfunction  
Event Description
It was reported that on (b)(6) 2023, a 12mm amplatzer vascular plug ii was chosen for treatment of an endo leak in a transcatheter aortic valve implantation (tavi) case.During preparation, the device was properly prepared per instructions for use (ifu).During procedure, it was noted that the device didn't open.The device was removed form the patient prior to release.The procedure was terminated.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient was reported recovering.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
An event of device deformity was reported.The investigation confirmed the device met functional and visual specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.It was noted that an unknown size delivery system was used which could have contributed to the reported event.However, it could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER VASCULAR PLUG II
Type of Device
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18536493
MDR Text Key333349548
Report Number2135147-2024-00287
Device Sequence Number1
Product Code KRD
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K071699
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9-AVP2-012
Device Lot Number8721711
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-