• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number HL-390
Device Problems Crack (1135); Material Separation (1562); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2023
Event Type  malfunction  
Event Description
It was reported that the reservoir was cracked, and missing water tank cap.Patient involvement unknown, and no patient injury or clinical affects was reported.
 
Manufacturer Narrative
D5.Operator of device is unknown; no information has been provided to date.Device evaluation: one device was received with a broken front cover, missing water tank cap, missing standoffs behind liquid crystal display (lcd), cracked water tank cover.The reported problem was verified by visual inspection.The most probable but unconfirmed root cause was over tightening of the screw.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced water tank cover, printed circuit board (pcb), water tank cap, front cover.Performed preventative maintenance (pm), changed o-rings, reservoir gasket and calibrated.The device passed all functional and delivery tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18538582
MDR Text Key333481235
Report Number2183161-2024-00069
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberHL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2023
Date Manufacturer Received12/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-