SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. MEDEX TRANSTAR KITS; PRESSURE MONITORING Back to Search Results
Catalog Number MX9505T
Device Problem Failure to Calibrate (2440)
Patient Problem Insufficient Information (4580)
Event Date 12/12/2023
Event Type  malfunction  
Event Description
It was reported that the nurse replaced the arterial pressure sensor at 9:00 on (b)(6) 2023 and failed to perform a zero calibration several times.The monitor prompted that the arterial pressure was high and could not be zeroed.It was then replaced with a new pressure sensor and was working normally.There was patient involvement and had patient harm/adverse event reported.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Device evaluation: a review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.No trend of confirmed complaints in relation with this issue was identified.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDEX TRANSTAR KITS
Type of Device
PRESSURE MONITORING
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
carretera miguel alemán km21.7
apodaca nuevo leon
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18539446
MDR Text Key333225660
Report Number9616567-2024-00014
Device Sequence Number1
Product Code DPT
UDI-Device Identifier10351688503992
UDI-Public(01)10351688503992(17)250919(10)4323171
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K942377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMX9505T
Device Lot Number4323171
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-