• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BRAEMAR MANUFACTURING, LLC C6 MCOT PPM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BRAEMAR MANUFACTURING, LLC C6 MCOT PPM Back to Search Results
Model Number 700-0718-01
Device Problems Difficult to Remove (1528); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported by the patient's daughter that the sensor charger adapter could not be removed from the wall socket.A replacement was ordered.There was no harm noted.
 
Manufacturer Narrative
It was reported that the duel charge adapter got stuck in the wall outlet.The device was not returned for investigation.Engineering evaluation was unable to be performed as the device was not returned for evaluation.The failure description aligns with a known failure realted to jdsp-18928 dual usb power adaptor housing separation and philips am&d is further investigating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C6 MCOT PPM
Type of Device
C6 MCOT PPM
Manufacturer (Section D)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer (Section G)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer Contact
beverly okoh
1285 corporate center drive
suite 150
eagan, MN 55121
MDR Report Key18539451
MDR Text Key333214350
Report Number2133409-2024-00002
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number700-0718-01
Was Device Available for Evaluation? No
Date Manufacturer Received04/09/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age88 YR
Patient SexFemale
-
-