Intuitive surgical, inc.(isi) did receive the maryland bipolar forceps (mbf) instrument to perform failure analysis.The reported complaint was not confirmed by failure analysis.The instrument was placed and driven on an in-house system.The maryland bipolar forceps (mbf) instrument passed the recognition, engagement, electrical continuity and energy delivery tests.The instrument moved intuitively with full range of motion in all directions.The grips opened and closed properly.The instrument was further inspected and found charring and localized melting at the grip base between the grips and on the outer surface of one bipolar yaw pulley, resulting in an exposed electrode.Thermal damage is attributed to carbonized tissue on the grips of this bipolar instrument creating a conductive path during use.Thermal damage can also result due to inadvertent energy application from a monopolar instrument.
|