Catalog Number M003WE0300150
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2023
Event Type  malfunction  
Manufacturer Narrative
H3 other text : the device is not available to the manufacturer.
 
Event Description
It was reported that during an endovascular procedure when delivering the subject stent to the target site, the surgeon felt very strong resistance at the bend on the way to the target site.After removing the guidewire from the patient's body, it was found that the dual-tapered tip of the subject stent was broken.The procedure was completed successfully by replacing the subject device with a different stent.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WINGSPAN 3.0MM X 15MM - CE
Type of Device
STENT, INTRACRANIAL NEUROVASCULAR
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key18542500
MDR Text Key333355303
Report Number3008881809-2024-00020
Device Sequence Number1
Product Code NJE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
H050001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberM003WE0300150
Device Lot Number24113446
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDE CATHETER (UNKNOWN)
-
-