The manufacturer was contacted in reference to a simplygo mini device with complaints of the device overheating and a burning smell.There is no allegation of patient harm or serious injury.There was no medical intervention specified.The device was visually inspected/evaluated by the manufacturer.The manufacturer could not confirm the complaint; however, during inspection there were alarms that occurred when running under high load operation and the device did not produce.The manufacturer concluded this could be due to the device operating environment or ventilation obstruction.An internal inspection was also completed and there were no signs of burnt parts.The sieve canister was replaced along with the compressor inlet filter.Due to an internal leak, the sheave 02 side assembly was also replaced.After inspection and running test, the device has been confirmed to be working properly.
|