A facility reported that the mayfield skull clamp (a1059) came loose during use: during the assembly of the skull clamp, the patient was positioned and during the calibration of the neuro-navigator spheres, the skull clamp disengaged, not keeping the patient in position.Patient was already under anesthesia; however there was no injury or significant surgical delay.Another skull clamp was used to complete the procedure.
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10 the mayfield skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation of the returned device showed a sticking swivel lock assembly and a residue buildup was present.Some worn off internal parts must be replaced to adjust the swivel lock assembly.To resolve the issues, the internal parts to adjust the unit according to the manufacturer were replaced and the swivel lock assembly was cleaned.After completing the reassembly, the skull clamp was subjected to a successful function test along with general cleaning and maintenance performed.Root cause - evaluation found no device deficiencies that would have contributed to the reported complaint.Probable root cause is improper or suboptimal positioning of the skull clamp on the patient.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|