MAUDE Adverse Event Report: SYNTHES GMBH CABLE CUTTER W/TRIGGER HANDLE; CUTTER, WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH CABLE CUTTER W/TRIGGER HANDLE; CUTTER, WIRE Back to Search Results
Catalog Number 03.221.006
Device Problems Break (1069); Device Slipped (1584); Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2024
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: reporter is a j&j employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on january 4, 2024, the orthopedic cable cutter was found to be broken in the set upon inspection after decontamination.No issues occurred during a case.The silver trigger grip was found to be loose and did not engage the cutting mechanism.This report involves one cable cutter w/trigger handle.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Per the lot# t106502 cable cutter w/trigger handle is assigned to part# 03.221.006, which is an assembly comprised of trigger handle component 03.221.007 & cable cutter blade assembly part# 03.221.008.Only was received the component trigger handle component 03.221.007 and visual analysis revealed that pusher is fallen apart and missing.Broken condition was not observed in the physicals device received.A dimensional inspection for the cable cutter w/trigger handle was not performed due to post manufacturing damage.A functional test was performed and the device was not work as expected, the silver trigger grip was found to be loose and did not engage the mechanism due to missing component of the pusher.The complaint condition was able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the cable cutter w/trigger handle would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: g1 mfg h4: device history: part number: 03.221.006.Lot number: t106502.Manufacturing site: tuttlingen.Release to warehouse date: 20-nov-2014.A review of the device history records was performed for the finished device lot number, and no non-conformances were identified.The raw material certificate was reviewed and the used material was according to the specification of the device.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CABLE CUTTER W/TRIGGER HANDLE
Type of Device
CUTTER, WIRE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
TUTTLINGEN
unter hasslen 5
tuttlingen CO 78532
GM   78532
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18545887
MDR Text Key333329558
Report Number8030965-2024-01100
Device Sequence Number1
Product Code HXZ
UDI-Device Identifier10886982076694
UDI-Public(01)10886982076694
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.221.006
Device Lot NumberT106502
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-