Catalog Number ASD37A
Device Problems Break (1069); Entrapment of Device (1212); Difficult to Remove (1528)
Patient Problem Retroperitoneal Hemorrhage (4483)
Event Date 01/09/2024
Event Type  Injury  
Manufacturer Narrative
The gore® cardioform septal occluder instructions for use states: adverse events associated with the use of the occluder may include but are not limited to: perforation of a cardiovascular structure by the device.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
It was reported to gore a 37mm gore® cardioform septal occluder was selected to treat an atrial septal defect.After the device was deployed and locked it was reported that it appeared that a portion of the right atrial disc was jutting into the superior vena cava.The physician tried to advance a balloon to try and move the right atrial disc into the correct position; however, this was unsuccessful.The physician then tried to retrieve the device but as the right disc was being brought back into the catheter the retrieval cord broke.The physician then tried to snare the device but was unsuccessful as it appeared as though the left atrial disc was lodge in the inferior vena cava causing what appeared to be a perforation of the inferior vena cava.Further attempts to remove the device were stopped as the patient became unstable and was taken to surgery.It was reported there was a retro-peritoneal bleed due to the injury.A long open abdominal approach was needed to remove the device and treat the peritoneal area.The patient remains in the intensive care unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® CARDIOFORM ASD OCCLUDER
Type of Device
TRANSCATHETER, SEPTAL OCCLUDER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
kathy titus
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key18547058
MDR Text Key333283558
Report Number2017233-2024-04555
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00733132636501
UDI-Public00733132636501
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberASD37A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/09/2024
Date Device Manufactured05/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age1 YR
Patient SexMale
Patient Weight9 KG
-
-