• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION WAVEWRITER ALPHA; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION WAVEWRITER ALPHA; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-1232
Device Problems Use of Device Problem (1670); Power Problem (3010); Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2023
Event Type  Injury  
Event Description
It was reported that following a non-device related surgery, the patients ipg was unable to be turned on and was unable to connect with the remote control or clinicians programmer.It was also noted that the patients right lead had migrated.The patient underwent a revision procedure wherein the ipg and lead were replaced.The patient was doing well postoperatively and unable to returned explanted device due to facility policy.Per the physicians implant manual, electrocautery is a known source of high voltage transient signal and warns against the use of electrocautery.
 
Manufacturer Narrative
B3: exact date unknown, event occurred in 03/23/2023.Additional suspect medical device component involved in the event: product family: scs-linear eads.Upn: m365sc2317700.Model: sc-2218-50 / sc-2317-70.Serial: (b)(6).Batch: 5073897 / 5089876.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVEWRITER ALPHA
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key18547147
MDR Text Key333284078
Report Number3006630150-2024-00149
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729985099
UDI-Public08714729985099
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/13/2023
Device Model NumberSC-1232
Device Catalogue NumberSC-1232
Device Lot Number517153
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age75 YR
Patient SexMale
Patient Weight102 KG
-
-