• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PNEUPAC PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PNEUPAC PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RES) Back to Search Results
Catalog Number P310NUS
Device Problem Failure to Cycle (1142)
Patient Problem Insufficient Information (4580)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3: date of event and d4: serial number are unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was stated that the device was not cycling properly.There was unknown patient involvement and unknown patient harm/adverse event reported.
 
Manufacturer Narrative
Additional information: d4: serial number and h4: manufacture date.H3 - other: device not returned for investigation.H6 - evaluation codes: updated.Device evaluation: no product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.The most probable cause is a faulty input manifold assembly.The device would stay in a continuous flow manner.Therefore, draining o2.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC PLUS
Type of Device
VENTILATOR, EMERGENCY, POWERED (RES)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18547532
MDR Text Key333286383
Report Number3012307300-2024-00325
Device Sequence Number1
Product Code BTL
UDI-Device Identifier35019315107263
UDI-Public35019315107263
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberP310NUS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/16/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-