The customer reported to olympus, the bronchovideoscope had a channel leakage.The issue was found during a unknown diagnostic procedure was completed with a similar device.There were no reports of patient harm or impact associated with this event.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was found: the connecting tube had coating peeling (1mm2 or more).This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned, and an evaluation was completed.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, ) it is likely physical stress was applied to insertion section, stress from inferior storage environment (in direct sunlight, at high temperatures, in high humidity or exposed to x-rays and/or ultraviolet-rays, etc.), and chemical stress from reprocessing not recommended by instructions for use (ifu) caused the coating to peel.However, a definitive root cause could not be determined.The following information is stated in the instructions for use (ifu): 3.2 "inspection of the endoscope" 5.Inspect the external surface of the entire insertion tube including the bending section and the distal end for dents, bulges, swelling, peeling, scratching, holes, sagging, transformation, bends, adhesion of foreign body, dropout of parts, any protruding objects or other irregularities.Olympus will continue to monitor field performance for this device.
|