Model Number 5610004 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/15/2023 |
Event Type
malfunction
|
Event Description
|
Circuits are inoperative: cracks in tubing, kinked and/or having holes.1 event occurred during use.Pt required bagging until circuit was replaced.1 event discovered during setup.Prior to pt use.
|
|
Manufacturer Narrative
|
The being cracked or kinked would disrupt the patient therapy.The clinician would need to replace the circuit during procedure causing an interruption in patient therapy.
|
|
Manufacturer Narrative
|
The being cracked or kinked would disrupt the patient therapy.The clinician would need to replace the circuit during procedure causing an interruption in patient therapy.The investigation confirmed the defect of damaged components based on the provided customer photos.The most likely root cause of the observed defect was identified to be the manufacturing and molding of the devices.A quality alert (24-036) has been issued to the manufacturing site as a result of these findings.
|
|
Event Description
|
Circuits are inoperative: cracks in tubing, kinked and/or having holes.1 event occurred during use.Pt required bagging until circuit was replaced.1 event discovered during setup.Prior to pt use.
|
|
Search Alerts/Recalls
|