Model Number NCEUP3012X |
Device Problems
Deflation Problem (1149); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/26/2023 |
Event Type
malfunction
|
Event Description
|
An attempt was made to use an nc euphora balloon to treat a lesion.The device was inspected with no issues.Negative prep was not performed.It was reported that balloon deflation difficulties occurred.The device was slow to deflate at the lesion site.No patient injury reported.
|
|
Manufacturer Narrative
|
Product analysis: the device was returned for analysis.No deformation was evident to the proximal balloon bond/inflation lumen.The balloon folds returned intact.Deformation was evident to the distal tip.The balloon passed negative prep.The balloon was inflated to 12 atms, and deflation testing was completed without issue.No other deformation evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Initial reporter name and phone number provided.Annex d code.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|