• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U125
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2023
Event Type  malfunction  
Event Description
It was reported that there was oversensing of noise on the right ventricular (rv) lead of this cardiac resynchronization therapy pacemaker (crt-p) system.The rv lead was not manufactured by boston scientific.Technical services (ts) analyzed the presenting electrogram (egm) and found that this resulted in multiple inappropriately stored ventricular events and approximately 2.5 seconds of pacing inhibition during one of them.Further testing in-clinic was advised.No adverse patient effects were reported.Currently, this crt-p remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18552276
MDR Text Key333357697
Report Number2124215-2024-02924
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559389
UDI-Public00802526559389
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number723189
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexMale
-
-