SIEMENS HEALTHCARE DIAGNOSTICS INC. CLINITEK STATUS+ Back to Search Results
Catalog Number 10379675
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2023
Event Type  malfunction  
Manufacturer Narrative
Siemens requested that the customer return their instrument as well as the power cord and batteries in use at the time of the event for investigation.Siemens has shown due diligence in requesting the return of the customer's instrument.However, siemens has not received any additional update on the return.Without the requested product, the investigation could not proceed.The cause of this event is unknown.It should be noted that if the customer was using the status connect platform, batteries are not recommended.Per the instrument manual: "caution do not use batteries in the analyzer, if you attach the analyzer to a clinitek status connector.Leaving the batteries in the battery compartment may corrode the batteries.".
 
Event Description
The customer reported that their status+ instrument produced visible smoke and a burnt smell.The customer stated that they were using both batteries and a power cord at the time of the event.The customer was advised to immediately unplug the unit and remove it from service.There is no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD.
northern road
chilton industrial estate
sudbury, suffolk CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
MDR Report Key18552287
MDR Text Key333357766
Report Number3002637618-2024-00003
Device Sequence Number1
Product Code JIL
UDI-Device Identifier00630414574639
UDI-Public00630414574639
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-