PHILIPS NORTH AMERICA LLC HEART START XL; DEFIBRILLATOR Back to Search Results
Model Number M4735A
Device Problem Failure to Discharge (1169)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2024
Event Type  malfunction  
Manufacturer Narrative
H3 other text : remote support provided.
 
Event Description
This report is based on information provided by philips service representative and has been investigated by the philips complaint handling team.Philips received a complaint on the heartstart xl indicating that the device cannot discharge in synchronous mode.The customer worked with the service representative.It was conveyed that the customer connects the lead wire when using synchronous mode, and r-wave can be detected before discharge.Based on the information available and the cause of the reported problem was user error.The reported problem was not confirmed.Serviceable ¿ product back in service ¿ no further action required: the device was operational after setup information was conveyed to the user.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART START XL
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
tanya deschmidt
1001 murry ridge lane
ste a
murrysville, PA 15668
9789415572
MDR Report Key18553052
MDR Text Key333367347
Report Number3030677-2024-00282
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K021453
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM4735A
Device Catalogue NumberM4735A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-