BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number 5C4482S
Device Problems Fluid/Blood Leak (1250); Fracture (1260); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a minicap transfer set had a fractured valve (twist clamp) and it did not close properly which resulted in a leak.This was observed during use of the device for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
B5: the patient was treated with 1 g of vancomycin and 100 mg of amikacin ip (intraperitoneally).No disinfectant products were used on the transfer set.The transfer set was replaced.B7: catheter implanted in september 2023.H10: the actual device was not available; however five (5) photographs of the sample were provided for evaluation.The photographs were visually inspected and a broken occluder leg beneath the twist clamp was observed causing a fluid flow through the set with the twist clamp closed.The reported condition was verified.The cause of the condition could not be determined; however, exposure to chemical such as foreign solvents, dyes, or cleaning agents that damage the transfer set materials.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18554130
MDR Text Key333383377
Report Number1416980-2024-00118
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412476261
UDI-Public(01)00085412476261
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C4482S
Device Lot NumberH23F15032
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/29/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age23 YR
Patient SexFemale
-
-