Reportable based on device analysis completed on 12-jan-2024.It was reported that the balloon failed to deflate.A 15mmx3.50mm wolverine balloon was selected to treat lesion located in the mildly tortuosity and moderately calcified left anterior descending artery (lad).During the procedure, when balloon insertion and inflation were attempted, inflation was not successful.The procedure was completed using another wolverine device and no patient complications were reported.However, return device analysis revealed pinholes on the balloon material.
|
Device evaluated by manufacturer: the returned product consisted of the wolverine cb mr, ous 15mmx3.50mm device.Visual, tactile, microscopic and functional analysis was performed on the device.A detailed microscopic examination of the balloon material identified two pinholes when inflated to rated burst pressure.One was 1mm distal from the distal markerband and the other was 1mm proximal from the proximal markerband.For functional analysis, the device was attached to an encore inflation device which was verified before and after use using the druck gauge.An attempt was then made to inflate the balloon to 12atm as per, wolverine ifu, using an inflation aid attached however, pressure was unable to be maintained as inflation liquid was observed leaking from pinholes.
|