• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 195274
Device Problem Device Alarm System (1012)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) observed that the red level sensor would not recognize fluid when placed on thin-walled side of testing device.The fsr replaced the red level sensor and performed release testing.The unit operated to manufacturer's specifications.During laboratory evaluation, the product surveillance technician (pst) observed that when the level sensor was placed on the thick-walled side of a reservoir testing simulator, the sensor functioned as intended and alerted when there was no liquid in front of the sensor.When tested on the thin-walled side of the reservoir testing simulator, whether there was liquid or no liquid in front of the sensor, it was alarming.It was determined that the red level sensor did not operate as intended.
 
Event Description
It was reported that the level sensor was falsely alarming.There were no reported adverse consequences to the patient.No other details regarding the nature of this event were provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key18557123
MDR Text Key333510000
Report Number1828100-2024-00028
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000496
UDI-Public(01)00886799000496(11)230810
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number195274
Device Catalogue Number195274
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/10/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-