U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2023
Event Type  malfunction  
Manufacturer Narrative
B3: date of event and d5: operator of device are unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the power does not turn on.The event occurred before patient use.
 
Manufacturer Narrative
H3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received.No visual observations were noted.Per functional testing, the main unit did not turn on when the power switch was turned on.The complaint was confirmed.The root cause was a defective ac power cord and printed circuit board (pcb).It was unknown what caused the defect.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced ac power switch and pcb.The device passed all functional and delivery tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
3350 granada ave n, suite 100
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18557885
MDR Text Key333460657
Report Number2183161-2024-00086
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public(01)50695085820909(11)161004
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberHL-90JP
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/09/2024
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-