DEPUY SPINE INC CONFIDENCE SPINAL CEMENT SYSTEM CONFIDENCE PLUS KIT SPINAL CEMENT SYSTEM 11CC; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Catalog Number 283910000
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2024
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: initial reporter is a synthes employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from taiwan reports an event as follows: it was reported that on (b)(6).2024 during an unknown procedure, the pipeline of the cement system in question was leaking.There was no reported adverse patient impact.No further information is available to report.This report is for a confidence spinal cement system confidence plus kit spinal cement system 11cc.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample and review of the photo/video evidence provided revealed that the flexible tube of the hydraulic pump was found cracked.The observed condition has in consequence that the device fails to contain fluid properly and leaks during use.A dimensional inspection was not performed since it was not applicable to the complaint condition.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.A functional test was performed connecting the hydraulic pump flexible tube to the cement reservoir cap and it was performed correctly.The connectors were connected and disconnected with no issue perceived and the handle of the hydraulic pump work as intended.However, due to the flexible tube has damage the fluid leak off.The complaint condition was able to be replicated.The overall complaint was confirmed as the observed condition of the confidence spinal cmt sys, 11c would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Dimensional inspection: n/a.Device history: a manufacturing record evaluation was performed for the finished device.Product code: 283910000; lot no: 377798.It was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 13-jul-2023; manufacturing site:jabil le locle; expiry date:30-jun-2025; cement number: product code :182901001 / batch number : 4137376.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIDENCE SPINAL CEMENT SYSTEM CONFIDENCE PLUS KIT SPINAL CEMENT SYSTEM 11CC
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
JABIL LE LOCLE
chemin blanc 36-38
le locle 2400
SZ   2400
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key18562567
MDR Text Key333459467
Report Number1526439-2024-00682
Device Sequence Number1
Product Code NDN
UDI-Device Identifier10705034209623
UDI-Public(01)10705034209623
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K060300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number283910000
Device Lot Number377798
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2023
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-