MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OPTICROSS HD; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OPTICROSS HD; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number 8668
Device Problems Unable to Obtain Readings (1516); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2023
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 4jan2024.It was reported that visualization issues occurred.The opticross hd imaging catheter was selected for ultrasound examination of the target lesion.During preparation, the screen was black and no image was shown.The procedure was completed with another of the same device.There were no patient complications reported and the patient condition following the procedure was stable.However, device analysis revealed the imaging window was twisted and the tip was stuck with the guidewire.
 
Manufacturer Narrative
The device was returned for analysis.Visual inspection revealed the sheath and imaging window were kinked, the imaging window and imaging core were twisted, and the tip was stuck with the floppy part of the guidewire.Microscopic inspection revealed the guidewire exit port and tip were in good condition.Impedance testing showed an electrical open at the proximal end of the catheter between 140-150 cm from the distal housing which is related to breakage of the coaxial cable.Functional testing could not be performed due to the condition in which the device was returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS HD
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18563342
MDR Text Key333477887
Report Number2124215-2024-03280
Device Sequence Number1
Product Code OBJ
UDI-Device Identifier08714729960768
UDI-Public08714729960768
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K173284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8668
Device Catalogue Number8668
Device Lot Number0031811288
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/15/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
Patient Weight80 KG
-
-