AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H180J
Device Problems Pressure Problem (3012); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported no insufflation on the evis exera ii gastrointestinal videoscope.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was identified: the nozzle of the insufflation channel was clogged by dark rubbery material.There were no reports of patient or user harm associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the evaluation found the reported malfunction was confirmed.A full technical evaluation was completed, and the results showed that there was air/water flow issues.The nozzle of the insufflation channel was clogged by dark rubbery material likely due to insufficient cleaning.Additionally, the charged coupled device (ccd) glue was worn-out.The connecting tube was buckled.The bending angulation was out of standard values.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over nine years since the subject device was manufactured.Based on the results of the investigation, the root cause of the reportable malfunction could not be determined.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: ifu states the detection method in evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 preparation and inspection ifu states the preventative measures in evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories) olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18564592
MDR Text Key333490674
Report Number9610595-2024-01458
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04953170307539
UDI-Public04953170307539
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K051645
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-H180J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-