BD MEDICAL - DIABETES CARE - HOLDREGE, NE / 68949 SYRINGE 0.5ML 31GA 8MM 10 BAG 500 WAL; Syringe, piston
|
Back to Search Results |
|
Model Number 328509 |
Device Problems
Leak/Splash (1354); No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
Consumer reported she could not draw her insulin to take injection.Stated, some of the insulin was leaking from the barrel.Stated, it happened last week but could not remember the exact date.1 syringe affected.Lot: 3156431.Catalog: 328509.Date of event: unknown.Sample: yes cl.
|
|
Manufacturer Narrative
|
Correction to h6, device code.Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
|
Search Alerts/Recalls
|
|
|