Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 12/29/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).D10: (b)(4).Item name g7 pps ltd acet shell 54f.Lot # j7614041.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the acetabular impactor threads fractured off inside of the implanted cup.The tip of the impactor was irretrievable inside the threads of the polar hole of the implant and was left in the implant and subsequently the patient.A new impactor was used with a head ball impactor tip for final impaction, the cup could not be removed and thus a new cup was not necessary.There is no additional information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified the threaded tip of the device fractured.The device is covered in bio-debris.No further evaluation can be made from the provided pictures.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no update to the prior event description provided.
|
|