Model Number V173 |
Device Problems
Over-Sensing (1438); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problems
Loss of consciousness (2418); Syncope/Fainting (4411)
|
Event Date 01/17/2024 |
Event Type
Injury
|
Event Description
|
It was reported that, this patient exhibited multiple syncopal episodes.Upon review, this cardiac resynchronization therapy pacemaker (crt-p) was found to be in safety mode and presented oversensing.Subsequently, this crt-p was explanted and replaced.No additional adverse patient effects were reported.
|
|
Event Description
|
It was reported that, this patient exhibited multiple syncopal episodes.Upon review, this cardiac resynchronization therapy pacemaker (crt-p) was found to be in safety mode and presented oversensing.Subsequently, this crt-p was explanted and replaced.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.The system resets occurred during a telemetry session and while communicating with the latitude remote monitoring system and caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
|
|
Search Alerts/Recalls
|