Model Number A22001A
Device Problems Loose or Intermittent Connection (1371); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.In addition to the reportable findings documented in b5, non-reportable findings are as follows; the label of the eyepiece was worn, the outer tube was from 3rd party repair, and there was no image.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.E1/establishment name: hulunbeier people's hospital (hulunbeier league people's hospital (added due to character limitation in respective field.).
 
Event Description
The customer reported to olympus, the telescope's image was blurry.The issue was found during preparation for use of a therapeutic urological surgery.The device was returned for evaluation.During the device evaluation, the following reportable malfunctions were found: the distal end was burnt and worn, and the blue ring of the eyepiece was loose.There was no delay or patient harm, nor was the patient under anesthesia.The procedure was able to be successfully completed using the same device.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 12°, 4 MM
Type of Device
HYSTEROSCOPE, GYNECOLOGY
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18572245
MDR Text Key333615658
Report Number9610773-2024-00237
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020848
UDI-Public04042761020848
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K897003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22001A
Device Lot Number659468
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2024
Was the Report Sent to FDA? No
Date Manufacturer Received01/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
VISERA PRO VIDEO SYSTEM CENTER.
-
-