COOK INC ULTRATHANE COPE NEPHROURETEROSTOMY SET; FAD STENT, URETERAL Back to Search Results
Model Number N/A
Device Problems Difficult to Remove (1528); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Event Description
It was reported that the stiffener of two ultrathane cope nephroureterostomy set separated during a percutaneous nephrolithotomy procedure (pcnl).Resistance was felt during attempted removal of the stiffener from the stent.The stiffener was removed halfway when it separated.As result, the entire stent was removed from the patient.Another of the same device was used, and the same incident occurred again.The procedure was completed successfully with a work-around performed by the physician.The physician used a 1:1 ratio of contrast and saline during the procedure and suggested that could be a possible reason for the difficult removal.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Manufacturer Narrative
G4: pma/510(k) #: k171603.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE COPE NEPHROURETEROSTOMY SET
Type of Device
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18573409
MDR Text Key333624974
Report Number1820334-2024-00113
Device Sequence Number1
Product Code FAD
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberULT10.2-10.2-26-NUCL-B-RH
Was Device Available for Evaluation? No
Date Manufacturer Received01/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-