EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number PXVP2284AT3
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that viscous fluid was found in the reservoir of a truwave vamp during device set up.Hospital staff noticed that a ring in the syringe was left behind from the fluid when pushing the plunger up and back.No patient involvement.
 
Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Manufacturer Narrative
A product evaluation of one single dpt vamp plus kit was completed.The reported event of fluid in tubing and vamp was not able to be confirmed.No visible presence of fluid was observed from the kit during pre decontamination examination.Kit was wiped with ipa during decontamination and further examined in product evaluation lab.Microscopic examination of returned kit after decontamination did not detect any presence of fluid inside vamp system and pressure tubing.A device history record review was completed and documented that device met all specifications upon distribution.Further investigation was not required since the reported event was not confirmed and no product defects were detected during product evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR  
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key18574561
MDR Text Key333697827
Report Number2015691-2024-00559
Device Sequence Number1
Product Code DXO
UDI-Device Identifier07460691950436
UDI-Public(01)07460691950436(17)241114(11)221115(10)64673581
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K925638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberPXVP2284AT3
Device Lot Number64673581
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-